The Q-Exactive Focus is a benchtop quadrupole Orbitrap system based on the combination of Hyberbolic Quadrupole technology combined with Orbitrap analyzer technique, designed for routine/small molecules applications. Implementation of a Quadrupole mass filter in front of the Curved ion trap (C-Trap) allows for precursor ion selection and hence for MS/MS and SIM scan modes besides the ability of full MS mode. PRM and SIM are also implemented in possible scan modes. In addition the high sensitivity RF-lens ion source, with the implementation of enhanced FT transient processing, provides scan speed and resolving power. Q-Exactive Focus is a very robust and very sensitive benchtop quadrupole high resolution mass spectromere perfectly combining precise quantification and high performance qualitative work in the instrument. This instrument is an Orbitrap-based FT-MS system providing high resolution, high mass accuracy and high sensitivity at an affordable price.
Mass range:
50 < m/z < 3000
Sensitivity:
Full MS: 500 fg Buspirone on column S/N 100:1
SIM: 50 fg Buspirone on column S/N 100:1
Resolution: at m/z 200 (scan rate)
17,500 (12 Hz)
35,000 (7 Hz)
70,000 (3 Hz)
Mass accuracy: < 1 ppm RMS error with internal calibration
Quadrupole Isolation: Stepless precursor isolation from 0,4 amu to full mass range
Scan modes: Full Scan, MS/MS, SIM, SIM-MSX, PRM, AIF, vDIA (not for USA)
Polarity switching: One full positive and negative scan cycle under 1 sec at a resolution Setting of 35,000 (at m/z 200)
HCD fragmentation:
NCE
CE
The Q-Exactive Focus system consists of a quadrupole mass filter fully integrated with an
Orbitrap Mass Spectrometer. Additionally included items:
— H-ESI II probe,
— Chemyx Fusion 100 Syringe pump with full data system control.
In terms of the above description the Swedish University of Agricultural Sciences, aim to purchase a High Resolution Masspectrometer (HRMS) of typ Q-Exactive Orbitrap LC-MS from Thermo Electron Sweden AB by negotiated procedure without prior announcement with reference to Chapter 6. Section 14 paragraph 2 of LOU.